公司大多数产品均可提供包括产品材质、产品型号、产品功能、产品包装等全方位定制服务。具体请咨询客服人员。
我们通过十几年的医疗和健康产品出口服务,积累了丰富的行业经验,可根据客户需求,通过提供专业和个性化服务满足全球不同客户对个性化及品牌化订制医疗产品的需求。
Model: HIP-EA3
01. DPS(动态压力系统)确保实时闭塞压力。
02.特别原创双时尚,密集和正常的营养模式
03.在工作期间不要更改预设数据,以免误操作,提高进料安全性
04. USB用于更新软件和记录历史数据
Model No. Parameters |
Enteral Feeding Pump |
Flow rate range(ml/h) | 1~400, step by 1ml/h |
Accuracy of flow rate | ±10%, with dedicated IV set |
volume limit(ml) | 0-9999, step by 1 ml |
Accuracy of infusion volume | ±10%, with dedicated IV set |
Prime rate and accuracy(ml/h) | 1000 ±10% |
Flush rate and accuracy(ml/h) | 1000 ±10% |
One-time flush volume(ml) | 40±10% |
Max infusion pressure | ≥0.1 MPa |
Occlusion alarm value(pressure) | 50KPa±10KPa |
IV set | Refer to the recommended IV set list |
Bubble sensitivity | Detestability of air bubble≥50uL |
Alarming | Occlusion, Battery exhausted |
Indicating functions | Bubble in line, VTBI completion, Flush completion, Low battery, Malfunction(ER series),Pause overtime, Power interruption, Infusion mode, Working state |
Noise | With equipment operating on a wooden table in the maximum flow rate, and in the distance of 1 meter in every direction, the noise level is below 50dB. |
Power | AC:100V-240V 50 Hz /60Hz |
Internal power | After fully charged, internal DC11.1V rechargeable Li battery can run over 5 hours at intermediate rate |
Weight | 2.5kg |
Classification | Type BF, Class II continuous volumetric feeding pump,IPX1(sealed drip-proof device) |
Size | 170mm(W)×155mm(L)×180mm(H) |
Optional Accessory | Fluid warmer, Dedicated disposable nutrition IV set |